首页 / 产品 / ELISA试剂盒 / 小鼠ELISA试剂盒
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of mouse IL-17 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 84 | ||||
Range % | 80-92 | |||||
1:2 | Average % | 90 | ||||
Range % | 85-100 | |||||
1:4 | Average % | 96 | ||||
Range % | 92-99 | |||||
1:8 | Average % | 97 | ||||
Range % | 92-105 |
The recovery of mouse IL-17 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 92 | 89-96 | ||||
EDTA plasma (n=4) | 97 | 94-103 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/ml | OD1 | OD2 | Average | Corrected | ||
5000 | 2.702 | 2.604 | 2.653 | 2.494 | ||
2500 | 2.055 | 1.967 | 2.011 | 1.852 | ||
1250 | 1.386 | 1.418 | 1.402 | 1.243 | ||
625 | 0.875 | 0.829 | 0.852 | 0.694 | ||
312 | 0.537 | 0.521 | 0.529 | 0.370 | ||
156 | 0.351 | 0.353 | 0.352 | 0.193 | ||
78 | 0.246 | 0.234 | 0.240 | 0.081 | ||
0 | 0.162 | 0.155 | 0.159 |
×