Intra-assay Precision (Precision within an assay): CV%<8% | |
Three samples of known concentration were tested twenty times on one plate to assess. | |
Inter-assay Precision (Precision between assays): CV%<10% | |
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of mouse IL-1β in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | |||
Sample | Serum(n=4) | ||
1:1 | Average % | 88 | |
Range % | 82-95 | ||
1:2 | Average % | 93 | |
Range % | 88-98 | ||
1:4 | Average % | 96 | |
Range % | 94-102 | ||
1:8 | Average % | 94 | |
Range % | 90-100 |
The recovery of mouse IL-1β spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | |||
Sample Type | Average % Recovery | Range | |
Serum (n=5) | 94 | 87-99 | |
EDTA plasma (n=4) | 93 | 88-98 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/ml | OD1 | OD2 | Average | Corrected | ||
2000 | 2.758 | 2.641 | 2.700 | 2.586 | ||
1000 | 1.438 | 1.587 | 1.513 | 1.399 | ||
500 | 0.833 | 0.869 | 0.851 | 0.737 | ||
250 | 0.474 | 0.477 | 0.476 | 0.362 | ||
125 | 0.350 | 0.375 | 0.363 | 0.249 | ||
62.5 | 0.267 | 0.253 | 0.260 | 0.146 | ||
31.25 | 0.165 | 0.186 | 0.176 | 0.062 | ||
0 | 0.119 | 0.108 | 0.114 |
A micro ELISA plate --- The 96-well plate has been pre-coated with anti-mouse IL-1β This dismountable microplate can be divided into 12 x 8 strip plates.
Two vials lyophilized standard --- Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
Biotin-labeled IL-1β antibody (100 x concentrate) 1 x 120 μl ---Act as the detection antibody.
HRP-avidin (100 x concentrate) 1 x 120 μl --- Bind to the detection antibody and react with the TMB substrate to make the solution chromogenic.
Biotin-antibody Diluent 1 x 15 ml ---Dilute the high concentration Biotin-antibody to an appropriate working solution.
HRP-avidin Diluent 1 x 15 ml ---Dilute the high concentration HRP-avidin solution to an appropriate solution.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
An incubator can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×