Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of mouse IL-2 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 86 | ||||
Range % | 80-92 | |||||
1:2 | Average % | 95 | ||||
Range % | 91-100 | |||||
1:4 | Average % | 96 | ||||
Range % | 90-101 | |||||
1:8 | Average % | 92 | ||||
Range % | 86-98 |
The recovery of mouse IL-2 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 95 | 89-98 | ||||
EDTA plasma (n=4) | 96 | 94-99 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/ml | OD1 | OD2 | Average | Corrected | ||
2000 | 2.151 | 2.234 | 2.192 | 2.078 | ||
1000 | 1.525 | 1.647 | 1.586 | 1.471 | ||
500 | 1.085 | 1.028 | 1.057 | 0.942 | ||
250 | 0.718 | 0.729 | 0.723 | 0.609 | ||
125 | 0.436 | 0.421 | 0.428 | 0.314 | ||
62.5 | 0.298 | 0.283 | 0.291 | 0.176 | ||
31.25 | 0.200 | 0.194 | 0.197 | 0.082 | ||
0 | 0.114 | 0.115 | 0.115 |
A micro ELISA plate ---The 96-well plate has been pre-coated with an anti-mouse IL-2 antibody. This dismountable microplate can be divided into 12 x 8 strip plates.
Two vials lyophilized standard ---Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
One vial Biotin-labeled IL-2 antibody (100 x concentrate) (120 μl/bottle) ---Act as the detection antibody.
One vial HRP-avidin (100 x concentrate) (120 μl/bottle) ---Bind to the detection antibody and react with the TMB substrate to make the solution chromogenic.
One vial Biotin-antibody Diluent (15 ml/bottle) ---Dilute the Biotin-antibody.
One vial HRP-avidin Diluent (15 ml/bottle) ---Dilute the HRP-avidin solution.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
An incubator can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×