首页 / 产品 / ELISA试剂盒 / 人ELISA试剂盒
Intra-assay Precision (Precision within an assay): CV%<15% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<15% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
ng/ml | OD1 | OD2 | Average | Corrected | ||
3600 | 2.578 | 2.634 | 2.606 | 2.606 | ||
1800 | 1.563 | 1.667 | 1.615 | 1.615 | ||
900 | 0.834 | 0.828 | 0.831 | 0.831 | ||
450 | 0.441 | 0.429 | 0.435 | 0.435 | ||
150 | 0.192 | 0.201 | 0.197 | 0.197 | ||
0 | 0.078 | 0.083 | 0.081 | 0.081 |
A micro ELISA plate ---The 96-well plate has been pre-coated with an anti-human sTfR antibody.
Six vials standard (5 ml/bottle) ---Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
One vial HRP-conjugated sTfR antibody (6 ml/bottle) ---Bind to the C-Peptide in the samples or standards and react with the substrate to make the solution chromogenic.
Substrate A (1 x 7 ml)
Substrate B (1 x 7 ml)
Four Adhesive Strips (For 96 wells) ---Cover the microplate when incubation.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 600 nm or 630 nm.
An incubator can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×