首页 / 产品 / ELISA试剂盒 / 人ELISA试剂盒
Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of human IL-1Ra in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 94 | ||||
Range % | 85-102 | |||||
1:2 | Average % | 90 | ||||
Range % | 86-100 | |||||
1:4 | Average % | 89 | ||||
Range % | 82-95 | |||||
1:8 | Average % | 95 | ||||
Range % | 87-105 |
The recovery of human IL-1Ra spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 97 | 93-105 | ||||
EDTA plasma (n=4) | 100 | 94-108 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
pg/ml | OD1 | OD2 | Average | Corrected | ||
1000 | 2.503 | 2.482 | 2.493 | 2.413 | ||
500 | 1.713 | 1.723 | 1.718 | 1.638 | ||
250 | 1.035 | 1.017 | 1.026 | 0.946 | ||
125 | 0.604 | 0.615 | 0.610 | 0.530 | ||
62.5 | 0.337 | 0.319 | 0.328 | 0.248 | ||
31.25 | 0.201 | 0.206 | 0.204 | 0.124 | ||
15.6 | 0.154 | 0.162 | 0.158 | 0.078 | ||
0 | 0.081 | 0.078 | 0.080 |
A micro ELISA plate ---The 96-well plate has been pre-coated with an anti-human IL1RN ANTIBODY. This dismountable microplate can be divided into 12 x 8 strip plates.
Two vials lyophilized standard ---Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
One vial Biotin-labeled IL1RN antibody (100 x concentrate) (120 μl/bottle) ---Act as the detection antibody.
One vial HRP-avidin (100 x concentrate) (120 μl/bottle) ---Bind to the detection antibody and react with the TMB substrate to make the solution chromogenic.
One vial Biotin-antibody Diluent (15 ml/bottle) ---Dilute the Biotin-antibody.
One vial HRP-avidin Diluent (15 ml/bottle) ---Dilute the HRP-avidin solution.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
An incubator can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×