Intra-assay Precision (Precision within an assay): CV%<8% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<10% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of mouse BMG/β2-MG in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 88 | ||||
Range % | 83-93 | |||||
1:2 | Average % | 92 | ||||
Range % | 85-98 | |||||
1:4 | Average % | 86 | ||||
Range % | 82-91 | |||||
1:8 | Average % | 93 | ||||
Range % | 85-99 |
The recovery of mouse BMG/β2-MG spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 93 | 87-98 | ||||
EDTA plasma (n=4) | 99 | 92-106 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
μg/ml | OD1 | OD2 | Average | |||
10 | 0.219 | 0.230 | 0.224 | |||
5 | 0.305 | 0.317 | 0.311 | |||
2.5 | 0.445 | 0.428 | 0.436 | |||
1.25 | 0.538 | 0.572 | 0.555 | |||
0.625 | 0.815 | 0.821 | 0.818 | |||
0.312 | 1.099 | 1.053 | 1.076 | |||
0.156 | 1.503 | 1.484 | 1.494 | |||
0 | 2.235 | 2.125 | 2.180 |
×