Intra-assay Precision (Precision within an assay): CV%<15% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<15% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
ng/ml | OD1 | OD2 | Average | Corrected | ||
0 | 0.125 | 0.131 | 0.128 | 0.128 | ||
12.5 | 0.287 | 0.269 | 0.278 | 0.278 | ||
37.5 | 0.436 | 0.428 | 0.432 | 0.432 | ||
75 | 0.697 | 0.689 | 0.693 | 0.693 | ||
150 | 1.173 | 1.221 | 1.197 | 1.197 | ||
300 | 2.049 | 2.053 | 2.051 | 2.051 |
A micro ELISA plate --- The 96-well plate has been pre-coated with anti-human hepcidin antibody.
Six vials standard (0.5 ml/bottle)--- Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
One vial HRP-conjugated hepcidin antibody (6 ml/bottle) --- Bind to the hepcidin in the samples or standards and react with the substrate to make the solution chromogenic.
One vial Wash Buffer (20x concentrate) (15 ml/bottle) --- Wash away unbound or free substances.
One vial Substrate A (7ml/bottle) --- Mix with substrate B and then interact with TMB to elicit a chromogenic reaction.
One vial Substrate B (7ml/bottle) --- Mix with substrate A and interact with TMB to elicit a chromogenic reaction.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
An incubator can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×