Intra-assay Precision (Precision within an assay): CV%<15% | ||||||
Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
Inter-assay Precision (Precision between assays): CV%<20% | ||||||
Three samples of known concentration were tested in twenty assays to assess. |
To assess the linearity of the assay, samples were spiked with high concentrations of human CA15-3 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
Sample | Serum(n=4) | |||||
1:1 | Average % | 90 | ||||
Range % | 86-99 | |||||
1:2 | Average % | 97 | ||||
Range % | 91-105 | |||||
1:4 | Average % | 95 | ||||
Range % | 89-100 | |||||
1:8 | Average % | 92 | ||||
Range % | 88-98 |
The recovery of human CA15-3 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
Sample Type | Average % Recovery | Range | ||||
Serum (n=5) | 97 | 93-99 | ||||
EDTA plasma (n=4) | 96 | 90-101 |
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||
U/ml | OD1 | OD2 | Average | |||
0 | 0.109 | 0.114 | 0.112 | |||
15 | 0.187 | 0.196 | 0.192 | |||
40 | 0.425 | 0.401 | 0.413 | |||
100 | 0.964 | 0.975 | 0.970 | |||
250 | 1.937 | 1.836 | 1.887 |
A micro ELISA plate --- The 96-well plate has been pre-coated with an anti-human CA15-3 antibody.
Five vials standard (0.5 ml/bottle)--- Dilute a bottle of the standard at dilution series, read the OD values, and then draw a standard curve.
One vial HRP-conjugated CA15-3 antibody (6 ml/bottle) --- Bind to the CEA in the samples or standards and react with the substrate to make the solution chromogenic.
One vial Sample Diluent (11ml/bottle) ---Dilute the sample solution.
One vial Wash Buffer (20x concentrate) (15 ml/bottle) --- Wash away unbound or free substances.
One vial Substrate A (7ml/bottle) --- Mix with substrate B and then interact with TMB to elicit a chromogenic reaction.
A microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at603 nm or 630 nm.
An incubator can provide stable incubation conditions up to 37°C±5°C.
Centrifuge
Vortex
Squirt bottle, manifold dispenser, or automated microplate washer
Absorbent paper for blotting the microtiter plate
×